DescriptionThe TE 79 Multi-Axis Tribometer is for friction and wear testing under moderate load
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
The TE 79 Multi-Axis Tribometer is for friction and wear testing under moderate loads in pin or ball on rotating disc or reciprocating plate configurations. In pin on disc mode the machine complies with ASTM G 99 and DIN 50 324. In both modes, the indexing capability allows tests to be performed in accordance ASTM G132. which requires indexation of the pin, so that it is always presented with a fresh abrasive surface.
Contact Configurations: | Ball on Flat |
Pin on Flat | |
Customised Specimens | |
Normal Load: | 0.1 to 100 N |
Load Measurement: | Databeam (strain gauge transducer) |
Friction Force Range: | 0 to 50 N |
Friction Measurement: | Databeam (strain gauge transducer) |
Humidity Sensor: | 10 to 90% RH |
Drive Motor: | Brushless d.c. servo motor – 200 W |
Load Motor: | Electric Linear Actuator, 12V dc, 100N |
Indexing Motor: | Electric Linear Actuator, 12V dc, 100N |
Interface: | USB Serial Link Interface Module |
Software | COMPEND 2000 |
TE 79/P Indexing Pin on Disc Module | |
Contact Configurations: | Ball on Disc |
Pin on Disc | |
Disc Diameter: | 100 mm |
Track Radius: | 0 to 40 mm |
Traverse Speed: | 10 mm/min |
Rotation Speed: | 0 to 300 rpm |
Sliding Speed: | up to 1 m/s |
TE 79/R Indexing Reciprocating Module | |
Contact Configurations: | Ball on Plate |
Plate on Plate | |
Plate on Hemisphere | |
Maximum Sliding Axis Speed: | 10 mm/s |
Maximum Sliding Stroke: | 50 mm |
Maximum Traverse Axis Speed: | 1 mm/s |
Maximum Traverse Stroke: | 30 mm |
Temperature Range: | ambient to 100°C |
Dwell (time delay): | User selected in seconds up to 8 hours |
Temperature Sensor: | J-type thermocouple |
Heating Power: | 150 W |
TE 79/H Non-indexing High Temperature Pin on Disc Module | |
Disc Diameter: | 75 mm |
Track Radius: | 0 to 35 mm |
Temperature: | Ambient to 500 °C |
TE 79/R/C Peltier Cooler | |
Minimum Temperature: | -15°C (ambient water cooled) |
Minimum Temperature: | -30°C (chiller water/glycol cooled) |
RE 79/R/C Laboratory Chiller | |
Working Fluid: | 50:50 Water/Glycol |
Fluid Temperature: | -35°C |
Controlled Parameters | Sliding Axis Position (TE 79/R) |
RPM (TE 79/P) | |
Sliding Axis Speed (TE 79/R) | |
Traverse Position (TE 79/P and TE 79/R) | |
Traverse Speed (TE 79/P and TE 79/R) | |
Temperature (TE 79/R) | |
Dwell Period | |
テスト期間 | |
Measured Parameters | Sliding Axis Position (TE 79/R) |
Traverse Position (TE 79/P and TE 79/R) | |
Humidity | |
Ambient Temperature | |
Temperature (TE 79/R) | |
Friction | |
Friction Coefficient | |
Services | |
Electricity: | 220/240 V, single phase, 50 Hz, 720 W |
110/120 V, single phase, 60 Hz, 720 W | |
Installation | |
Bench-mounting machine: | 570 mm x 600 mm x 600 mm high, 40 kg |
Bench-mounting controller: | 530 mm x 530 mm x 240 mm high, 20 kg |
Packing Specifications: | 0.59 m3, GW 120 kg, NW 70 kg |