Technical parameters:1. Combustion time :0——9999 seconds can be set arbitrarily, the standard uses 3
電子メール:info@standard-groups.com
Technical parameters:
1. Combustion time :0——9999 seconds can be set arbitrarily, the standard uses 30 seconds;
2. Flame and blazing time timer :0——9999 seconds, accuracy :+0.1 seconds;
3. Combustion test chamber size :660mmx660mmx1065mm;
4. Fuel :5%(V/V) methanol and 95%(V/V) ethanol mixture;
5. Flame height :150mm——180mm, adjustable;
6. Fuel consumption :2.55±0.15mL/min.
7. Alcohol blowtorch: hanging type;
8. Automatic timing of burning time;
9. Automatic timing of open flame and blazing time
10. Flame height adjustable.
11. Steel plate electrostatic spraying box, with ventilation device
12. Pure copper fire nozzle, equipped with preheating tank, preheating operation is simple.
13, equipped with a flame adjustment device, can test the flame height according to the standard requirements,
14. Equipped with firehead manual handle, moving the firehead is safe and convenient,
15. The inner box is pure black (the standard requires the inner box to be dark)
16. Fuel :5%(V/V) methanol and 95%(V/V) ethanol mixture (customer's own);
17. Weight :60KG Packing volume :0.85 cubic meters