Cleveland Open Cup Auto Flash Point Analyser, ISO2592 - Qinsun
専門検査機器メーカー
全国問い合わせホットライン:+86 021-64208466
製品

Cleveland Open Cup Auto Flash Point Analyser, ISO2592

Products:The instrument complies with ISO2592 "petroleum products of flash and fire assay (Cle

WhatsApp : +86 13816217984

Email : info@qinsun-lab.com

コンテンツ

Products:

The instrument complies with ISO2592  "petroleum products of flash and fire assay (Cleveland open cup)," the other standard methods, the use of microcomputer technology control, automated work process, color touch-screen operation, high-speed thermal the printer automatically print the results, and has failed self-test function, and it could be used in petroleum, chemicals, power, environmental protection, railways, research to do flash point testing.

Model: STD-ISO 2592 

Standard: ISO2592

Main Features:

1.7-inch color touch-screen LCD display, English interactive display interface;

2. PID control algorithms automatically adjust the heating curve according to standards;

3. Equipped with automatic detection chip atmospheric pressure to measure atmospheric pressure.

4. A big size data storage function, can store 300 results for any time historical record;

5. Automatic error correction system, automatic stop when temperature is alarm;

6. Mixing, ignition, detection, alarm, cooling, print is automatically controlled;

7. The testing process automatically lidopenning, ignition, alarm, cooling and printing;

8. Platinum alloy heating wire ignition, stable and reliable, high precision;

9. Can be connected to a laptop via USB interface communication, compatible withWINDOWSXP /WINDOWS7 system, network management.

10. Multilingual language operation.

11 ,Warming, openings, fire, alarm, cooling, printing is automatically controlled by system.

12. High-speed signal processors, high-precision analog / digital conversion circuit, stable instrument performance and test accuracy.

技術パラメータ:

1. Display Mode:7-inch color touch-screen LCD display
2. Measurement:Differential detection type thermocouple
3. Measuring Range:20 -400 ºC
4. Repeatability:±2ºC
5. Printer:High-speed thermal printer, 56 mm paper width
6. Ignition:Electronic ignition
7. Temperature Detection:Imported precision platinum resistance
8.Self-test function:Printer, ignition, solenoid valve, the lift arm is detected, the ignition sweep and so draw rod
9. Power Consumption:<700W
10. Use of the environment:10ºC-45ºC,30-80%  Temperature of 10 ºC -45 ºC, humidity 30-80%
11. Dimensions:450×380×290mm
12. Power :AC220V±10%
13. Weight:About 15 kg


免責事項: 上記のコンテンツは,業界内部者のみの参照およびコミュニケーションのためであり,その正確性または完全性を保証しません.関連法規および本ウェブサイトの規制により,関連品を購入するユニットまたは個人は,有効な資格および資格条件を取得する必要があります.


オンラインメッセージ

連絡先情報

会社電話

+86-21-6420 0566

営業時間

月曜日から金曜日

携帯電話:

13816217984

メール:

info@qinsun-lab.com

QRコード