Detailed Product DescriptionBias Voltage Range:0 ~ 500mvMeasuring Range:0.1~10000 Ng/μlTemperat
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Bias Voltage Range: | 0 ~ 500mv | Measuring Range: | 0.1~10000 Ng/μl |
---|---|---|---|
Temperature Control Range: | Room Temperature ~ 1000 ℃ | Temperature Control Accuracy: | ±1°C |
Measurement Accuracy: | Sample Concentration (ng/μl) 0.2 RSD(%)35 Sample Concentration (ng/μl) 1.0 RSD (%)10 Sample Concentration (ng/μl) 100 RSD(%)5 Sample Concentration (ng/μl)1000 RSD(%)2 | Gas Source Requirements: | General Nitrogen And General Oxygen |
Power Supply Voltage: | AC 220V±10% 50Hz±10% 3.5KW | Dimensions: | Host :410×350×75(mm) Temperature Control: 530×420×360(mm) Stirrer:290×270×360(mm) Injector:350×130×140(mm) |
Highlight: | Easy To Operate Chloride AnalyzerMicrocoulomb Sulfur Chloride Analyzer |
The Microcoulomb Sulfur Chloride Analyzer is designed and manufactured according to SH/T 0253. It applies microcoulomb analysis technology, uses oxidation method to oxidize the sample into titratable ions through the cracking furnace, and titrates in the titration cell. According to Faraday's law, calculate the sulfur content in the sample, which is suitable for light petroleum products with a boiling point of 40-310°C. Samples with a sulfur content ranging from 0.5 to 1000ppm, and samples with a sulfur content greater than 1000ppm should be measured after dilution. This instrument can also measure the content of chlorine.
Man-machine direct dialogue, easy to operate
The computer controls the whole process of analysis and data processing, displays the working status of the whole process, and saves or prints the parameters and results as required.
International popular circuits and imported components are used to reduce the noise of the instrument and improve the detection sensitivity.
It has the characteristics of stable and reliable performance, easy operation, high analysis precision and good repeatability.
Bias voltage range | 0 ~ 500mv |
Measuring range | 0.1~10000 ng/μl |
Temperature control range | room temperature ~ 1000 ℃ |
Temperature control accuracy | ±1°C |
measurement accuracy | Sample concentration (ng/μl) 0.2 RSD(%)35 Sample concentration (ng/μl) 1.0 RSD (%)10 Sample concentration (ng/μl) 100 RSD(%)5 Sample concentration (ng/μl)1000 RSD(%)2 |
Gas source requirements | general nitrogen and general oxygen |
Power supply voltage | AC 220V±10% 50Hz±10% |
Power | 3.5KW |
寸法すんぽう | Host :410×350×75(mm) Temperature control: 530×420×360(mm) Stirrer:290×270×360(mm) Injector:350×130×140(mm) |
免責事項: 上記のコンテンツは,業界内部者のみの参照およびコミュニケーションのためであり,その正確性または完全性を保証しません.関連法規および本ウェブサイトの規制により,関連品を購入するユニットまたは個人は,有効な資格および資格条件を取得する必要があります.
会社電話
+86-21-6420 0566
営業時間
月曜日から金曜日
携帯電話:
13816217984
メール:
info@qinsun-lab.com