Product introduction The guide wire friction tester is mainly used to measure the friction between
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Product introduction
The guide wire friction tester is mainly used to measure the friction between the guide
wire and the guide wire. This device is of great significance for research and practice in the
medical field because it can help doctors understand the movement characteristics of the
catheter and guide wire in the human body, so as to optimize surgical procedures and
improve treatment results.
Test principle
The sample is clamped between the two chuck of the fixture, the two chuck do relative
movement, and the puncture or opening force test will be carried out through the special
chuck. Through the force sensor located on the moving clamp head and the displacement
sensor built in the machine, the change of force value and displacement in the test process
is collected, and the corresponding force value data is obtained.
Test standard
YY/T1536-2017.ISO 11070
Technical parameter
1. Test range: 0-500N (50N, 100N, 200N, 1000N other ranges are optional)
2. Measurement accuracy: 0.5 level
3. Travel: 800mm
4. Test speed: 0.5-500mm/min stepless transmission
5. Displacement accuracy: 0.01mm
6. Sample quantity: 1 piece
7. Chuck size: 0-30mm
8. Shape: 450mm(L)× 500mm(B)× 1500mm(H)
9. Weight: 75kg